AltoStar® Automation System AM16

Key features

Entire workflow

  • The AltoStar®AM16 is CE-IVD marked
  • DNA and RNA purification in parallel with one-for-all chemistry AltoStar®Purification Kit 1.5 (CE-IVD)
  • Real-time PCR reaction setup by using the altona Diagnostics PCR kits (CE-IVD)
  • Automated cycler programming with AltoStar®Connect software (CE-IVD)
  • Expandable system for additional thermal cycler (CFX96™ Deep Well Dx System, IVD / Bio-Rad)

Process safety

  • Traceability: barcoded reagents for full process control
  • Ready-to-use reagents for consistent quality
  • Lot checking
  • Expiry date checking
  • Total Aspiration and Dispense Monitoring (TADM) – proven technology from Hamilton

High flexibility

  • All sample types in one run
  • Analyze up to four pathogens from a single sample
  • Combine up to eight real-time PCR assays on a 96 well plate
  • Scalable from 1–96 reactions, no minimum batch size
  • 500 µl and 1000 µl sample volume possible in one run
  • Sample carriers can be unloaded any time after sample processing

Any sample type

  • CE-IVD marked purification kit validated for whole blood, EDTA plasma, urine, stool and swabs in UTM
  • One purification kit for both DNA and RNA assays
  • One universal internal control for entire workflow (AltoStar® Internal Control 1.5, CE-IVD)

Convenience

  • Ready-to-use reagents for extraction and real-time PCR
  • Use of primary or secondary sample tubes
  • Freely selectable positions for samples
  • Load backup reagents and consumables
  • No refill or mixing of solutions necessary
  • Full access to eluates
  • User friendly control software AltoStar®Connect (see under AltoStar® product range)

Name Info Cat No Principal
AltoStar® Automation System AM16 Automatic Platform for robotic pipetting workstation used for sample purification and PCR setup controlled by the AltoStar® Connect software AM16 altona Diagnostics, Germany

Additionally included in package AltoStar® Molecular Diagnostic Workflow

Products AltoStar® Connect software
AltoStar® Plate Sealer
AltoStar® AM16 PC

Dimensions and weight

Dimensions (W x D x H) 1124 mm x 903 mm x 795 mm (transport size)
1010 mm (with loading tray)
1200 mm (with loading tray and carriers on loading tray)
Weight Approx. 140 kg

Operating data

Maximum power consumption ≤ 600VA
Voltage 115 VAC / 230 VAC ± 10 %
Delayed action fuse 115 V~: 6.3 A (T6.3AL250)
230 V~: 3.15 A (T3.15A250)
Installation category II
Pollution degree 2
Temperature range 15 °C to 30 °C / indoor use only
Relative humidity 15 % to 85 % (no condensation)
Noise level < 65dBA (regarding EN27779)
Altitude Up to 2000 m above sea level

Transportation and storage

Transportation Truckage company / AltoStar® AM16 instrument is
installed or relocated by a Hamilton trained
field service engineer.
Temperature range -25 °C to +70 °C / Indoor storage only
Relative humidity 10 % to 90 % (no condensation, indoors)

Reviews